August 2001 C-1
Appendix C: Scanner Specifications
Site Specifications
AC Power Requirements 1,750 watts peak
AC Frequency 50 or 60 Hz
AC Phase Single
Operating Environment Room temperature: 65 to 80° F (18 to 27° C)
Relative humidity: 20 to 70%
Ambient light: less than 1000 lux; no exposure to direct lighting
Altitude: 7000 feet (2134 m) maximum
Scanner Specifications
Dimensions and Weight Height: 15 in. (38,1 cm)
Width: 24 in. (61 cm)
Depth: 30 in. (76,2 cm)
Weight: 120 lbs (54,5 kg)
Footprint: 5 square feet
Operating and Service
Clearances
Rear: 1.5 feet (45,7 cm) for operation,
3 feet (91,4 cm) for service
Sides: 2 feet (60 cm) for operation and service
Front: 4 feet (122,9 cm) for operation and service
Scanner Location The scanner location relative to the host workstation is limited by the SCSI
interface cable length; if you are using a cable stocked by Kodak, this
distance is 2 meters or less. Longer cables are not supported for use with
the scanner. The scanner must be the only SCSI device attached to the
host computer.
The scanner should be located on a table that:
is stable
can support approximately 120 lbs (54,5 kg)
is at least 30 in. (76,2 cm) long x 30 in. (76,2 cm) wide
Power Consumption
and Thermal Load
Power Consumption: 425 watts maximum; 400 watts typical
Thermal Load: 1325 BTU operating; 1230 BTU standby
AC Voltage and Current
Draw
100 V AC/50 Hz 4.2 amps operating; 3.9 amps standby
100 V AC/60 Hz 4.1 amps operating; 3.85 amps standby
120 V AC/50 Hz 3.5 amps operating; 3.2 amps standby
200 V AC/60 Hz 2.0 amps operating; 1.9 amps standby
200 V AC/50 Hz 2.0 amps operating; 1.85 amps standby
230 V AC/60 Hz 1.8 amps operating; 1.7 amps standby
250 V AC/50 Hz 1.75 amps operating; 1.65 amps standby