Kodak HR 500 User Manual Page 45

  • Download
  • Add to my manuals
  • Print
  • Page
    / 118
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 44
Maintaining the Equipment
August 2001 4-5
2. Loosen the thumbscrews on the lamp cover and pull the cover straight out.
When you remove the cover, the safety switch is released, removing power to
the lamp socket.
CAUTION: Wait at least five minutes for the lamp to cool.
3. Push down the lamp release lever.
4. Remove the old lamp from the lamp socket.
5. Fully seat the new lamp in the socket; polarity is not important.
CAUTION: In the next step, to ensure correct alignment with the light bar,
it is extremely important to push the lamp to the back of the
lamp holder with the leads facing to the left. Improper
alignment affects the operation of the scanner.
6. With the socket leads facing left, reinstall the lamp in the lamp holder.
7. Reinstall the lamp cover and tighten the thumbscrews.
Power is reapplied to the lamp socket and the lamp should light.
8. Calibrate the scanner (see Calibrating the Scanner on page 3-4).
P192_0023GC
thumbscrews
lamp
lamp cover
socket
leads
Hot Surface
symbol
Page view 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 117 118

Comments to this Manuals

No comments